Benefits of a contralateral routing of signal device for unilateral Naída CI cochlear implant recipients
Abstract
PURPOSE : Many bilaterally deaf adults are only able to receive one cochlear implant (CI), resulting in suboptimal listening performance, especially in challenging listening environments. Adding a contralateral routing of signal (CROS) device to a unilateral CI is one possibility to alleviate these challenges. This study examined the benefit of such a CROS device.
METHODS : Thirteen adult subjects with at least 6 months of CI use, and no or limited benefit of a hearing instrument in the contralateral ear were included in the study. The perceived benefit of a CROS device in everyday listening environments was evaluated up to 1 year after initial fitting using several questionnaires. Speech intelligibility performance was determined using the French matrix sentence test in quiet and in two speech-in-noise setups and was followed for 3 months after CROS fitting.
RESULTS : Subjects indicated high satisfaction with the practical usability of the CROS device and long-term device retention was high. Perceived benefits in everyday listening environments were reported. Formal speech intelligibility tests revealed statistically significant median improvements of 6.93 dB SPL (Wilcoxon Z = 2.380, p = 0.017) in quiet and up to 8.00 dB SNR (Wilcoxon Z = 2.366, p = 0.018) in noise. These benefits were accessible immediately without a need for prolonged acclimatization.
CONCLUSIONS : Subjective satisfaction and device retention as well as speech intelligibility benefits in quiet and in noise prove the CROS device to be a valuable addition to a unilateral CI in cases of bilateral deafness where bilateral implantation is not an option.
Fichier principal
CROS_Paris_Rev_TextOnly.pdf (159.49 Ko)
Télécharger le fichier
Fig1_sized.pdf (188.62 Ko)
Télécharger le fichier
Fig2_sized.pdf (189.38 Ko)
Télécharger le fichier
Fig3_sized.pdf (173.93 Ko)
Télécharger le fichier
Fig4_sized.pdf (309.01 Ko)
Télécharger le fichier
modifiedAPS-SSD_Experience.pdf (615.67 Ko)
Télécharger le fichier
modifiedAPS-SSD_Satisfaction.pdf (630.6 Ko)
Télécharger le fichier
Origin | Files produced by the author(s) |
---|