Coexistence of circulating HBsAg and anti-HBs antibodies in chronic hepatitis B carriers is not a simple analytical artifact and does not influence HBsAg quantification

Résumé

Background : Presence at the same time of HBsAg and anti-HBs antibodies (HBsAg/Ab) is an entity sometimes encountered in chronic hepatitis B (CHB) carriers. Objectives : This study was designed to characterize such serological profiles and to assess the reliability of serological marker quantification by three commercially available assays in this setting. Study design : Among 2578 CHB identified patients, 129 (5%) had an HBsAg/Ab profile as determined by Abbott Architect. After exclusion of co-infections (HIV, HCV, HDV), HBV reactivation or HBIg treatment, 101 samples from 62 patients were tested for HBsAg and anti-HBs quantification using Architect, DiaSorin Liaison-XL and Roche Modular-Cobas. Influence of genotype and HBsAg variants was studied in 31 samples with HBV replication. Results : HBsAg detection was confirmed with the 3 techniques for 98% (n = 99) of the samples while the HBsAg/Ab profile was concordant between all techniques for 65% of them. The overall correlation between the 3 HBsAg quantification techniques was good (R2: 0.94–0.97). The median HBsAg concentration was comparable for the 99 samples whatever the used technique but a bias of −0.11 and 0.02 log IU/mL were noticed for DiaSorin and Roche compared to Abbott, respectively. Anti-HBs quantifications were poorly correlated between techniques with major discrepancies observed. Genotype and substitutions within the “a” determinant showed an impact on HBsAg quantification. Conclusions : The double HBsAg/Ab profile is not an analytical artifact and is confirmed on all commercially available techniques. While such profile does not influence HBsAg quantification, differences of HBsAg quantification were noticed according to HBV genotype or HBsAg variant.

Domaines

Virologie
Fichier principal
Vignette du fichier
Pancher_2015_Coexistence_of.pdf (174.8 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01102734 , version 1 (13-01-2015)

Identifiants

Citer

Marie Pancher, Nathalie Désiré, Yen Ngo, Sepideh Akhavan, Coralie Pallier, et al.. Coexistence of circulating HBsAg and anti-HBs antibodies in chronic hepatitis B carriers is not a simple analytical artifact and does not influence HBsAg quantification. Journal of Clinical Virology, 2015, 62, pp.32-37. ⟨10.1016/j.jcv.2014.11.015⟩. ⟨hal-01102734⟩
537 Consultations
1322 Téléchargements

Altmetric

Partager

Gmail Mastodon Facebook X LinkedIn More