MOSARIX: Multi-crystal spectrometer in the tender x-ray range at SOLEIL synchrotron - Sorbonne Université
Journal Articles Review of Scientific Instruments Year : 2024

MOSARIX: Multi-crystal spectrometer in the tender x-ray range at SOLEIL synchrotron

Denis Céolin

Abstract

We have built and commissioned a novel standalone multi-crystal x-ray spectrometer (MOSARIX) in the von Hamos configuration based on highly annealed pyrolytic graphite crystals. The spectrometer is optimized for the energy range of 2–5 keV, but this range can be extended up to 20 keV by using higher reflection orders. With its nine crystals and a Pilatus detector, MOSARIX achieves exceptional detection efficiency with good resolving power (better than 4000), opening the door to study small cross section phenomena and perform fast in situ measurements. The spectrometer operates under a He atmosphere, which provides a flexible sample environment for measurements in gas, liquid, and solid phases.
Fichier principal
Vignette du fichier
RSI24-AR-00097.pdf (636.47 Ko) Télécharger le fichier
Origin Files produced by the author(s)

Dates and versions

hal-04583638 , version 1 (22-05-2024)

Identifiers

Cite

Iyas Ismail, Roba Moussaoui, Régis Vacheresse, Tatiana Marchenko, Oksana Travnikova, et al.. MOSARIX: Multi-crystal spectrometer in the tender x-ray range at SOLEIL synchrotron. Review of Scientific Instruments, 2024, 95 (5), ⟨10.1063/5.0199230⟩. ⟨hal-04583638⟩
55 View
39 Download

Altmetric

Share

More